Beamline Information » Beamline by Number » BL15A1 

BL15A1

Multiple-Wavelength Anomalous Dispersive and Monochromatic Protein Crystallography

Biological Crystallography (BX, MX, PX) is a technique used to determine the 3-dimensional structures of biological molecules such as proteins, viruses and nucleic acids (RNA and DNA) to a resolution higher than ~5 Angstrom. From the studies of this structural information, the detailed mechanism of the biological functions carried out by those macromolecules in living cells and organisms can be elucidated. The x-rays deflected ("scattered") by a single crystal of biological molecules concentrate into sharp intense spots (crystal diffraction pattern), and the structure of the biological molecule can be determined by analysis of the intensities and positions of the diffraction spots.

►Contact Information
 
If you have any technical concerns, please contact us!
 
BL15A1 Spokesperson:
Jean, Yuch-Cheng (ycjean@nsrrc.org.tw, +886-3-578-0281 Ext: 7321)
 
BL15A1 Beamline Manager:
Chou, Chung-Kuang (ck.chou@nsrrc.org.tw, +886-3-578-0281 Ext: 7310)
 
 
►Technical Information
 
Operational (general use)
Yes, Since Jan, 2013.
%Time available (general use)
75%
Experiments
MAD, SAD, & Monochromatic
Source
14 Poles, 6.1 cm Period Length, 3.1 Tesla Superconducting Wiggler
Ring Current (mA)
360 mA
Average Beam Lifetime
Infinite Under NSRRC Top Up Conditions
Energy Range
5.6 ~ 15.5 keV (2.21 Å ~ 0.8 Å)
Energy Resolution
2 x 10E(-4)
Spot Size
FWHM: 0.49 mm (H) x 0.19 mm (V)
Optics
Vertically Collimating Premirror, LN2-Cooled Fixed-Exit Double Crystal Si(111) Monochromator, Toroidal Focusing Mirror
Flux
6 x 10E(11)
Flux Detailed
Flux Measured Through a 0.2 mm Round Aperture @ 12.65 keV
Wavelength Min (Å)
0.8
Wavelength Max (Å)
2.21
Monochromator
LN2-Cooled, Fixed-Exit Double Crystal Monochromator
Crystal Type
Si(111)
Mirrors
Rh Coated
Mirror Cutoff
15.5 keV
Special Capabilities
Automated Sample Mounting
Goniometer
Bruker MD2 Microdiffractometer
Pin Length
Manual Sample Mounting: No Restrictions; Automated Sample Mounting: 14 – 18 mm
Restrictions
Omega No Limit
Xtal-detector Distance
100 – 850 mm
Detector
Rayonix MX300HE CCD Area Detector
Detector Details
Inter-Frame Total Dead Time: 2s
2theta Range
Vertical Offset + 250 mm, - 100 mm
Horizontal Offset +/- 15 mm
Cryo Capability
Oxford Instrument CryoJet, ~110K @ Sample
Biohazard
BL-1
Software
BluIce Control Software, HKL/HKL2000, Strategy, CHOOCH
Hardware
3 2-CPU SGI Octane2s, 1 4-CPU SGI O350 Server, 1 Shared 12-CPU SGI Origin3800 Server, Shared 2.0 TB SGI Fiber Channel Raid System
Backup Devices
USB2.0 / IEEE1394 Portable Disk
Beamline Phone Number
886-3-5780281 Ext 1151
 
 
 